Trouble Shoo ng/guide for electrical issues with M2 MULTI-TIP chairs.
1. If only the le /last ba ery light is on.
2. If the 2 smaller lights at the arrow bu ons on the
hand control is on.
HMN a/s
NB! AS A FIRST COURSE OF ACTION
CONCERNING ELECTRICAL ISSUES, AL-
WAYS TRY TO CHARGE THE CHAIR
FOR 24 HOURS!
The chair is low on power and must be charged. The
chair will not elevate if only 1 ba ery light is on.
The emergency STOP has been ac vated. It must be
deac vated by holding down the emergency stop
bu on on the hand control un l the chair beeps once
and the 2 small diodes at the arrow keys go out.
3. All 4 lights at the top of the hand control is blink-
ing.
The chair's control box has received an error signal.
The chair must be reset. This is done by first pressing
the emergency stop bu on on the hand control un l
the emergency stop is ac vated, and then holding
the bu on next to the emergency stop bu on. First
the chair will beep 3-4 mes, then it will lower the
chair completely. When the chair is sta onary, it is
reset.
If the chair a er this is s ll blinking in all 4 top lights
on the hand control it means there is a
defect component.