Pole Mount Assemblies and Accessories
INS301
Integra® Pole Mount Sliding Bracket
INS400
Evolution™ Pole Mount Assembly with cm H
INS400CM Evolution™ Pole Mount Assembly with cm H
INS400L
Evolution™ Pole Mount Laser Level with bracket
INS410
Evolution™ Pole Mount Assembly with cm H
scale of -25 cm H
INS410CM Evolution™ Pole Mount Assembly with cm H
and Line Level
INS8903
Line Level Assembly (MR Safe)
Description
The Evolution™ Pole Mount As-
sembly, provides support and
alignment for the External CSF
Drainage and Monitoring Systems.
These systems are used for drain-
ing cerebrospinal fluid (CSF) from
an intraventricular catheter or
lumbar catheter to an external CSF
drainage system.
The Integra® Pole Mount is pro-
vided with a laser leveling device
and an optional line level. The laser
level is attached to the Integra Pole
Mount Assembly via a bracket. See
Figure 5 on page 6, for laser label
radiation information. The Pole
Mount Assembly with laser level is
intended for use with the Integra
External CSF Drainage Systems.
Pole Mount assemblies can be
used with following External CSF
Drainage and Monitoring Systems
INS8600, INS8601, NL850-8600,
NL850-8601,
INS1100,
INS9020 and INS9030.
Caution: Refer to Instructions
for Use (IFU) for the External CSF
Drainage System prior to using the
system or leveling the device.
O/-18 mm Hg and Line Level
2
Features and Benefits
The Pole Mount Assembly Incor-
porates:
• A fixed rail graduated in cm H
and/or mm Hg (+ or - 1.0 mm for
the full scale).
Standard Rail Scale:
INS400: -6 mm Hg to +23 mm Hg /
-9 cm H
INS400CM: -9 cm H
+32 cm H
Available with extended Scale:
INS410: -18 mm Hg to +23 mm Hg /
-25 cm H
INS410CM: -25 cm H
+32 cm H
• A sliding bracket which locks se-
curely into pole mount by tight-
ening adapter knob.
• A laser level, which slides into
place at the pole mount base and
is level with the zero reference
10-110,
point for easy alignment with
patient's anatomical reference
point.
• A line level (optional use, instead
of laser level) which attaches to
zero reference point at the pole
mount base for easy alignment
with patient's anatomical refer-
ence point.
O and mm Hg Rail with Laser Level and Line Level
2
O Rail, Laser Level and Line Level
2
O and mm Hg Rail with Laser Level and negative
2
O Rail and negative scale of -25 cm H
2
The pole mount assembly is easily
mounted to an IV pole by securely
tightening a single locking knob
located at the Pole Mount base.
O
The Pole Mount Assembly is de-
2
signed to be re-usable.
Indications
The Pole Mount Assembly provides
support and alignment for External
O to +32 cm H
O
2
2
CSF Drainage Systems. The Integra
O to
External Drainage and Monitoring
2
O
Systems allow for drainage and
2
monitoring of CSF from the later-
al ventricles of the brain and the
lumbar subarachnoid space in se-
O to +32 cm H
O
2
2
lected patients to reduce intracra-
O to
nial pressure (ICP), to monitor CSF,
2
O
to provide temporary drainage of
2
CSF in patients with infected CSF
shunts, and to monitor ICP.
Intended Purpose
Pole Mount Assemblies and Exter-
nal CSF Drainage and Monitoring
Systems are used in selected pa-
tients with severe head injury, sub-
arachnoid hemorrhage. Reyes syn-
drome or similar encephalopathies,
hydrocephalic Shunt infections, in-
tracranial hemorrhage and pre and/
or post-operative monitoring.