Application for measuring
fever
You can also use cyclotest® lady to
measure fever. Take the measurement
as described on Page 21.
On
average,
measured orally is 36.7 °C ± 0.5 °C.
You can use the guidelines below to
classify the level of fever:
37.5 °C to 38.0 °C: High temperature
38.0 °C to 39.0 °C: Moderate fever
Above 39.0 °C:
Temperatures measured rectally are
generally 0.5 °C higher than those
measured orally, while temperatures
measured under the arm are 0.5 °C
lower.
Warning: always consult a doctor if the
reading shows a high fever.
Error messages
The measured temperature is below
32.00 °C and is thus outside the
measurement range.
body
temperature
High fever
EN-28
Technical data
The measured temperature is above
42.99 °C and is thus outside the
measurement range.
Electronic fault in the device. If the
fault persists, please contact Customer
Service at UEBE Medical GmbH.
Technical data
Type
Maximum thermometer
Measurement
32.00 °C to 42.99 °C,
range
"Lo" display for Low (too
low) at temperature below
32.00 °C,
"Hi" display for High (too high)
at temperature above 42.99 °C
Measuring
± 0.10 °C between 35.50 °C
accuracy
and 42.00 °C at an ambient
temperature of 10 °C to 40 °C
and rel. air humidity of 30-85 %
Display
Liquid crystal display (LCD)
with four digits, smallest unit
for display 0.01 °C
End of
if temperature increase less
measurement
than 0.1 °C in 16 s.
Storage and
Temperature -10 °C - +60 °C
transportation
Rel. air humidity 25 % - 90 %
conditions
Battery
Button cell type LR41 or SR41,
1.55 V
IP rating:
IP 22: Protection from solid
foreign bodies with diameters of
12.5 mm and above; protection
against
EN-29
water
droplets