Herunterladen Diese Seite drucken

SRC-Statistikprotokoll

---------------------------
OPTI LION
SRC-Statistik
TT-MMM-JJ
Level 1
SRCID:XXXXXXX Verf:TTMMMJJ
S/N: XXXXX
Anzahl gemessen: 26
Anzahl ok : 26
Grenzwerte:
pH
7.100-7.150
Na
120.0 - 125.0 mmol/l
K
2.5 - 3.0 mmol/l
Datum
pH
08Aug
7.13
09Aug
7.12
10Aug
7.11
11Aug
7.10
12Aug
7.07
13Aug
7.08
14Aug
7.09
15Aug
7.11
16Aug
7.09
17Aug
7.08
19Aug
7.10
20Aug
7.11
21Aug
7.12
22Aug
7.13
23Aug
7.12
24Aug
7.11
25Aug
7.10
26Aug
7.10
27Aug
7.11
28Aug
7.11
29Aug
7.12
30Aug
7.13
31Aug
7.13
01Sep
7.11
02Sep
7.12
03Sep
7.11
---------------------------
MW
:
7.107
SD:
0.015
CV%:
0.21%
Gebrauchsanweisung – OPTI LION Elektrolyt-Analysator
HH:MM
Na
K
OK?
122.7
2.74
OK
122.5
2.72
OK
122.5
2.69
OK
122.8
2.71
OK
122.7
2.70
OK
122.8
2.71
OK
122.8
2.69
OK
122.5
2.71
OK
122.5
2.72
OK
122.6
2.71
OK
122.6
2.70
OK
122.5
2.69
OK
122.6
2.70
OK
122.5
2.71
OK
122.7
2.69
OK
122.8
2.68
OK
122.7
2.68
OK
122.5
2.70
OK
122.6
2.71
OK
122.6
2.72
OK
122.7
2.71
OK
122.8
2.73
OK
122.7
2.72
OK
122.7
2.74
OK
122.6
2.73
OK
122.5
2.72
OK
1.9
1.7
2.8%
3.0
ANHANG D - PROTOKOLLFORMATE
D-3