For further informations see PC software DME 4, menu bar
"Help".
To perform an output calibration you have to con-
nect a voltmeter respectively ammeter of sufficient
accuracy to the output terminals.
Current output / Voltage output
(Output A: x = 1, output B: x = 2, output C: x = 3, output
D: x = 4)
Variantes
Output
Brx01 Rx43
0 Ω
mA
open
output
(950685) (951360)
or
soldered
V output
sol-
open
dered
Calculation of resistors Rx45 and Rx46 for the full-scale
output currents Y2 in the range ≥ 1 to ≤ 20 mA:
Current output
1
Rx45 // Rx46 =
Y2 [mA]
0.99158 V
Y2
Rx45
Order No.
20
mA
open
–––
10
mA
open
–––
5
mA
open
–––
2.5 mA
2.7 kΩ 951 245
1
mA
3.3 kΩ 951 253
Calculation of resistors Rx34 and Rx44 for the full-scale
output currents Y2 in the range ≥ 1 to ≤ 10 V:
Voltage output
Rx34 // Rx44 = Y2 [V] · 27'229.4
Y2
Rx34
Order No.
10
V
270 kΩ
951 485
5
V
270 kΩ
951 485
2.5 V
68 kΩ
951 419
1
V
27 kΩ
951 360
The locations of the variable components on the output board
are shown in Figure 14.
Rx34
Rx44 Rx45
Rx46
27 kΩ
open variabel variabel
variabel vari- 0 Ω
never
abel
(950685) mind
or
soldered
1
27 kΩ
Rx46
Order No.
49.9 Ω
102 575
100
951 089
200
101 717
470
951 162
1.5 kΩ
951 211
Rx44
Order No.
open
–––
270 kΩ
951 485
open
–––
open
–––
R446
R445
R443
R343
Br301
Br401
R344
R444
R434
R334
D
C
Fig. 15. Location of the variable components.
8. Maintenance
No maintenance is required.
9. Dimensional drawing
M2.5 x 10
14
2
173.5
160
Fig. 15. EURAX DME 440, front plate width 14 TE.
R346
R246
R245
R345
R243
Br201
R244
R234
R143
B
Br101
R144
R134
A
7.45
55.88
14 TE
70.82
Fixing plane
R146
R145
41