6. Operation via menu functions
It is recommended to have the instrument calibrated if one or both of the
two factors (offset or span factor) have been changed.
EN
6.3
Logger
The "Logger" menu item is always displayed in the menu, even for instruments that
were not delivered with a data logger enabled.
6.3.1 Start / Stop
When the logger is started, the LOG symbol is displayed.
When the logger is started, the interval set for this logger process and the set logger
duration are displayed below stop.
If no value is entered for the logger duration, it will not be displayed and the instrument
will store values until the memory is full or the batteries are empty, whichever comes
first.
If the logger process is interrupted in an uncontrolled manner due
to empty batteries, the logger values are not lost as they are stored
immediately during the logger process.
After changing the battery, the instrument restarts, though the logger
process is not continued, it must be restarted.
Logged values are always measured values of the sensor and not filtered
values of the indication.
If the logger memory is full, the logger process is automatically stopped.
The logger file contains the following values:
Logger settings (once at the start)
■
Indicated pressure value if measuring interval = logger interval
■
Arithmetic mean value if measuring interval < logger interval
■
Peak Min./Max. if interval > sample rate
■
Temperature value
■
Set zero value
■
Density
■
40
WIKA operating instructions, model CPG1500