Possible causes of errors:
In case of an error in the program process, briefly turn off the balance and disconnect
from power supply. The weighing process must then restart from the beginning.
Fault
Display is not lit up.
The display does not change
when a sample is being loaded • Sample dish / dish holder is fitted incorrectly.
The weight display changes
constantly / the stability display
does not appear.
Incorrect measuring result
Measurement is taking too
long
Measurement is not
reproducible
Drying does not start
DAB-BA-e-1813
Possible cause
• The appliance is not switched on.
• The mains supply connection has been
interrupted (mains cable not plugged in/faulty).
• Power supply interrupted.
• Fuse has blown
• Sample dish has contact with windshield or
heating hood.
• Draught/air movement
• Table/floor vibrations
• Electromagnetic fields / static charging (choose
different location/switch off interfering device if
possible)
• Check adjustment
• No resetting to zero before loading the sample
• Incorrect shut-down criteria set
• Sample is not homogenous
• Drying time is too short
• Drying temperature too high (e.g. oxidation
sample material, boiling point of sample
exceeded)
• Temperature sensor soiled or defective
• Heating hood open
• The mains supply connection has been
interrupted (mains cable not plugged in/faulty).
49