Instruction Manual
3.4
Current sensor function in the case of SITRANS TK-H
Independently of the applied temperature, a parameterisable output current
between 4 and 20 mA can be set to test the output circuit.
3.5
Line compensation
Line compensation is necessary for the following measurements:
-
Two-wire resistance thermometers or resistance
-
Two-wire thermocouple with external Pt100 reference point
Line compensation is set by numerically specifying the measured line resistance
(sum of the foward and return conductors)