Calibration
The thermometer is initially calibrated at the time of manufacture. If the thermometer is used
according to the use instruction, periodic readjustment is not required. However, Welch Allyn
recommends checking calibration on an annual basis or whenever clinical accuracy of the
thermometer is in question. Procedures for checking calibration are outlined in the 9600 Plus
Calibration Tester (REF 01802-110) directions for use manual.
The above recommendations do not supersede the legal requirements. The user must always
comply with legal requirements for the control of the measurement, functionality, and accuracy
of the device which are required by the scope of relevant laws, directives or ordinances where
the device is used.
Manufacturing date is given by the LOT number located on front of the product. The first three
digits give the day of the year manufactured. The second two digits give the year of
manufacture. The last six digits represent the serial number of the device.
An example: LOT 27710 this product has been manufactured on the 277 day of the year 2010.
Product specifications
Displayed temperature range:
Operating ambient temperature range:
Display resolution:
Accuracy for displayed temperature range:
Clinical repeatability:
Long term storage ranges
Temperature:
Humidity:
Shock:
Measurement time:
Automatic «power down»:
Battery life:
Battery type:
Thermometer dimension:
Thermometer weight:
This infrared thermometer meets requirements established in ASTM Standard E 1965-98 (for the
thermometer system [thermometer with probe cover]). Full responsibility for the conformance of the product
to the standard is assumed by Kaz USA, Inc., 250 Turnpike Road, Southborough, MA 01772, USA.
ASTM laboratory accuracy requirements in the display range of 36 °C to 39 °C (96.8 °F to 102.2 °F) for infrared
thermometers is ± 0.2 °C (± 0.4 °F), whereas for mercury-in-glass and electronic thermometers, the
requirement per ASTM Standards E 667-86 and E 1112-86 is ± 0.1 °C (± 0.2 °F).
A clinical summary is available upon request.
24
All manuals and user guides at all-guides.com
20 –42.2 °C (68 – 108 °F)
10 – 40 °C (50 – 104 °F)
0.1 °C or °F
± 0.2 °C (35.5 – 42 °C) ( 95.9 – 107.6 °F)
± 0.3 °C (outside this temperature range)
± 0.14 °C (± 0.26 °F)
–20 °C to 50 °C (–4 °F to 122 °F)
95% non condensing
withstands drop of 3 feet
Initial start-up time: 3–4 seconds
Measurement time: 2–3 seconds
60 seconds
6 months/1000 measurements
2 x MN 1500 or 1.5 V AA (LR6)
152 mm x 44 mm x 33 mm (6˝ x 1.7˝ x 1.3˝)
100 g (3.6 oz) without batteries
This appliance conforms to the following standards:
DIN EN 60601-1: 3/96 «Medical electrical equipment» –
Part 1: General requirements for safety
DIN EN 12470-5: 2003 «Clinical thermometers» –
Part 5: Performance of infrared ear thermometers (with maximum device)
This product conforms to the provisions of the EC directive 93/42/EEC
(Medical Device Directive).
To protect the environment, dispose of empty batteries according to national
or local regulations.
W E L C H A L LY N A U S T R A L I A P T Y LT D
U N I T 5 3 8 S O U T H S T R E E T
R Y D A L M E R E , N S W 2 116
A U S T R A L I A
TYPE BF
EQUIPMENT
Internally Powered Equipment.
Continuous Operation.
Not protected against ingress of water.
Service
The Braun ThermoScan thermometer has a limited three year warranty. Should service be
required during or after the warranty period you must contact Welch Allyn at one of the
locations listed on the inside cover to obtain a return authorization number and shipping
address. Repackage the thermometer carefully in its original packaging or securely packed to
avoid damage during shipping. Include the original sales slip indicating the date of purchase, a
note describing the problem, your authorization number, and your return address. Send the
Braun ThermoScan thermometer prepaid and insured.
Note: Shipping address varies by country.
Answers to any questions regarding operation of the Braun ThermoScan thermometer may be
obtained by contacting Welch Allyn at one of the numbers listed on the inside cover.
Limited three year manufacturer's warranty
For Model PRO 4000
The PRO 4000 instrument is warranted by Kaz to be free from defects in material and
workmanship under normal use and service for a period of three years from the date of delivery
to the first professional user who purchases the instrument.
This warranty does not cover batteries, damage to the probe window, or damage to the
instrument caused by misuse, negligence or accident, and extends to only to the first
purchaser of the product. Additionally this warranty becomes void, if the thermometer is
operated with anything other than genuine Braun ThermoScan
Thermometer service or replacement will be carried out by Kaz subject to the terms of this
warranty. Loss or damage in return shipment to Kaz shall be at purchaser's risk. Kaz reserves
the right to replace a non-conforming product with the most comparable product currently
available. This product has a five year minimum expected lifetime.
0297
®