Herunterladen Diese Seite drucken

Hach si792 D Bedienungsanleitung Seite 61

Gelöster sauerstoff 2-draht -transmitter

Werbung

Tabelle 19 Busparameter des Umformerblocks (Fortsetzung)
Parameter
SECONDARY_VALUE_2
SECONDARY_VALUE_
UNIT_2
TEMP_SENSOR_TYPE
TEMP_WIRE_IMPEDANC
E
TEMP_SENSOR_CAL
Concentration/Saturation
PROCESS_PRESSURE_
UNIT
Description
Process temperature value
and status
Value
Status
Degree C or degree F.
Changes the unit of
temperature being displayed
and transmitted.
Type of temperature sensor.
The value entered must
correspond to the temp.
sensor being used.
Sets the wire impedance of
the temp. sensor. Typically 0
unless the wire of the sensor
gets too long
Desired temperature reading,
used for temperature
measurement calibration.
Selects the process pressure
measurement unit
Data
R/W Bytes
Range
type
R
4
DS_65
R
1
1001 = °C
R/W 2
uns16
1002 = °F
1000 = NTC30
R/W 2
uns16
1002 = NTC22
R/W 4
float
Default: 0
-10 – 10 K
R/W 4
float
Default: 0
1133 = kPA
R/W 2
uns16
1137 = BAR
1141 = PSI