Additional documentation for Operating Manual
Proline Prosonic Flow 92F
Information on "rupture disk" order option
Sensor housings with integrated rupture disks are optionally available.
#
Warning!
• Please note that the housing can no longer assume a secondary containment function if a rupture disk is used.
• Make sure that the function and operation of the rupture disk is not impeded through the installation. Triggering
overpressure in the housing is 10 to 15 bar. Take adequate precautions to ensure that no damage occurs, and risk
to human life is ruled out, if the rupture disk is triggered.
The position of the rupture disk is indicated by an adhesive label on top of the disk. If the rupture disk is triggered, the
adhesive label is damaged and can thus be visually monitored.
Additional sign regarding the position of the rupture disk (RUPTURE DISK)