1. Entity Parameters
P
= 166 mW (linear output curve)
0
C
= 15 µF
i
L
= 95 µH
i
1)
Ex-relevant electrical connection parameters
ATEX
The approved digital sensors with Memosens technology may be connected with the following
measuring cables
• CYK10
• or a Memosens measuring cable that is identical both in terms of design, device technology
and function
only to the ATEX-approved intrinsically safe sensor output circuit of the Mycom S CPM153-
G... or Liquiline M CM42 transmitter or alternatively to an ATEX-certified, intrinsically safe
Memosens sensor output that complies with the following maximum values. In particular, the
effective inner inductance and the capacitance of the approved, intrinsically safe sensor
output may not exceed the values below:
1)
1. Entity Parameters
U
= 5.1 V
0
I
= 130 mA
0
P
= 166 mW (linear output curve)
0
C
= 15 µF
i
L
= 95 µH
i
1)
Ex-relevant electrical connection parameters
FM / CSA and Nepsi
Pay attention to the associated Operating Instructions of your transmitter.
3
Incoming acceptance and product identification
3.1
Incoming acceptance
1.
Verify that the packaging is undamaged.
 Notify your supplier of any damage to the packaging.
Keep the damaged packaging until the matter has been settled.
Endress+Hauser
Incoming acceptance and product identification
1)
2. Entity Parameters
P
= 112 mW (trapezoid output curve)
0
C
= 14.1 µF
i
L
= 237.2 µH
i
2. Entity Parameters
U
= 5.04 V
0
I
= 80 mA
0
P
= 112 mW (trapezoid output curve)
0
C
= 14.1 µF
i
L
= 237.2 µH
i