If the instrument is placed on stand-by for about
10 minutes, it will automatically switch-off. When
the instrument is turned on again, the last measured
value will automatically be displayed.
Maintenance
The NycoCard
™
READER II has three replaceable
accessories, the white calibration device, (9) the pen
tip (14) and the pen ring (13). It is recommended
that these accessories are inspected regularly and
replaced if dirty or damaged.
Replace the pen tip: remove the pen tip, carefully snap
the tip off with your fingers. Replace the new pen tip.
Replace the pen ring: remove the pen ring, hold the
case shield with one hand and carefully wriggle the
ring off with the other hand. Replace the new pen ring.
Make sure it fits correctly.
Replace the calibration device: remove the device by
simultaneously pressing the small handle down and
the opposite side upwards. Place the new calibration
device.
• Be careful not to touch the optics of the
Reader pen.
• Any dirt or damages of the pen tip or
white calibration device may cause wrong
measure ments.
Replace the batteries:
Step 1. Use a screwdriver to open the battery lid (15).
Step 2. Remove the old batteries (16).
Step 3. Replace the batteries with three new batteries.
Step 4. Replace the battery lid. New batteries need to
be charged before using the instrument.
• Be sure that the batteries are inserted in the
right direction.
• When disposing the batteries, please follow
recommendations from the supplier.
Disposal of the NycoCard™ READER II
For correct disposal according to the Directive
2012/19/EU (WEEE), please contact your
local NycoCard
7
™
READER supplier.