PCE Instruments PCE-DFG NF 10K-ICA Bedienungsanleitung Seite 24

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

4.2
Settings
When you are in measurement mode, press the OK key to enter the settings screen which is
divided into 2 pages:
Page 1
In order to change settings, select the menu item with the arrow keys and confirm with the OK
key. The values can then be changed by means of the arrow keys. Then press "OK" to confirm
the settings or the Back key to discard.
Function
Display Unit
Force Area
Zero Tracking
Sampling Speed
Calibrate Grav
Alarm Upper
Description page 1
The display unit can be set to: „N", „kg", „lb" or „KPa"
The force area can be set to a value between 999.99cm
and is included in the calculation if the display unit selected is „kPA"
(important for the accuracy).
For zero tracking, you can set:
„Off", „0.1 N", „0.2 N", „0.3 N", „0.4 N", „0.5 N"
Values below the value set are automatically excluded before the
zero point stabilises. After stabilisation of the reading, the sampling
rate will be 1 x per second. Deviations from the measured value
which are below the set value are automatically excluded in order to
keep the displayed values.
You can set how many measurements are taken per second. A value
between 6 and 1600 Hz can be set here.
Note: The higher the sampling rate, the lower the accuracy will be.
Higher sampling rates are suitable for dynamic measurements
whereas lower sampling rates are more suitable for static and slow
measurements.
Enter the gravity at the place of calibration.
The upper alarm can be set to +/- 9999.9.
© PCE Instruments
21
Page 2
2
2
and 0.01cm