Herunterladen Diese Seite drucken

Ortholumm ML5/1 Bedienungsanleitung Seite 20

Therapeutisches led lichtgerät

Werbung

Anhang
GUIDANCE AND MANUFACTURERS DECLARATION – ELECTROMAGNETIC EMISSION
The (equipment or system) is intended for use in the electromagnetic environment specified below. The
customer or the user of the (equipment or system) should assure that it is in such an environment
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonics emis-
sions
IEC 61000-3-2
Voltage fluc-
tuations/flicker
emissions
EC 61000-3-3
Immunity test
Electrostatic
discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips,
short
20
Compliance
Electromagnetic environment - guidanc
GROUP 1
The (equipment or system) uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are un-
likely to cause any interference in nearby eletronic equipment.
Class B
The (equipment or system) is suitable for use in all establish-
ment, including domestic establishment and those directly
connected to the public low-voltage power supply network that
supplies building used for domestic purposes.
Not applicable
Not applicable
IEC 60601 test
Compliance level
level
±6 kV contact
±6 kV contact
±8 kV air
±8 kV air
±2 kV for power
±2 kV for power
supply lines
supply lines(!)
±1 kV for input/
±1 kV for input/out-
output lines
put lines
±1 kV differential
±1 kV differential
mode
mode
±2 kV common
±2 kV common
mode
mode
<5 % UT (>95 %
<5 % UT (>95 % dip
dip in UT) for 0,5
in UT) for
cycle
Electromagnetic environment-guidance
Floors should be wood, concrete or
ceramic tile. If floors are cover with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial or hospital environ-
ment
Mains power quality should be that a
typical commercial or hospital environ-
ment.
Mains power quality should be that a
typical commercial or hospital environ-
ment. If the