Current
Parallel connection
Stability
Internal resistance
- static
- dynamic
Recovery time
Ripple voltage
:«0.5
mV
-
r
For series connection according
to "master and slave" system
see under point 3e, chap. II. B.
: Max.
6 A for any value of the out-
put voltage.
: An arbitrary number of sets may
be connected in parallel.
For parallel connection according
to "master and slave" system
see under point 3f, chap. II. B.
: For mains voltage variations of
+ 10 % of the nominal value,
the
fluctuation of the output voltage is
« 0.03 % of the set value or < 2.5 mV.
Whichever is greater applies.
: < 10 mQ for slow changes from no
load to full load.
: For sinusoidal load variations
of
80 % from full load to full load and
a frequency of:
0.1 kHz < 0,010 Q
11
AREAS OOAD
Q
LOW
kHz
O0 07510
100
kHz < 0.100 2
250
kHz «0.100
Q
: < 50 usec for a square-wave load
variation of 80 % from full load to
full load.
By recovery time is understood the
time which elapses between the be-
ginning of the load variation and
the moment after which the deviation
of the direct voltage remains within
the limits determined by the static
internal resistance.
See fig. 6.
Ins.