9.
In case of overheating the thermal protection is triggered and the device will
cease to emit light. The display will indicate "Oh" and the fans will turn on for
1 minute. After cooling, the device will continue its normal operation, but stays
in Pause mode. Overheating can only happen in an emergency situation or
by accidental clogging of the cooling vents. For this reason the cooling vents
located on the bottom of the bleaching head must always be unobstructed in
order to allow air circulation.
VIII. DAILY CARE AND MAINTENANCE
1.
Disinfection of the PVC window of the bleaching head
For each patient, the emitting window is cleaned
with a cotton swab soaked in an alcohol solution.
2.
Cleaning the device:
To disinfect the device and its parts, spray disinfectant on a piece
of soft cloth or cotton and clean.
Do not use abrasives or solvents, as this may damage parts of the device!
IX. PREPARATION FOR WORK WITH PATIENTS
LED bleaching activation device BLANCONE ARCUS+ is an accessory to a medical device and is designed
to whiten hard dental tissues by photoactivation of a bleaching gel based on hydrogen or carbamide perox-
ide, intended for light activation, applied in dental practice (concentrations of H2O2 or their equivalent: 0.1
– 6% (cosmetic bleaching) or above 6% (medical bleaching).
1.
Before starting work with the bleaching device, insulate the patient's
non-calcified soft tissue, place a cover or protective napkin on the face,
apply protective UV cream and put on protective glasses. Bleaching gel
is applied to the teeth surface according to the directions for use outlined
by the gel manufacturer.
The patient should be under constant supervision by the dentist during
the bleaching process in order to avoid discomfort and to ensure all safe-
ty and technological procedures are applied.
2.
It is desirable to check the individual sensitivity of the patient. The device has two levels of intensity
(high HI and low LO) and the optimal conditions are selected by adjusting the distance of the device
to the patient's teeth.
ULTRA
Testing starts with a high intensity "Hi" and a distance of 5 cm and if the heat sensation is very strong,
the distance is increased to achieve an acceptable thermal sensation.
DUETTO
If the acceptable distance is more than 10 cm, it is better to switch to a mode with low intensity - "Lo"
HOME
NIGHT
(Low) and put the bleaching head closer to the teeth to obtain optimal distance.
DUETTO
ACTIVE
CLICK
ARCUS
LIPS
ARCUS
DUETTO
CARE
Dental LED bleaching activation unit - FIXED
EN
13