XRpad2 HWC-M Detectors User Manual
Table 15

Guidance and Manufacturer's Declaration of Electromagnetic Immunity

Guidance and Manufacturer's Declaration of Electromagnetic Immunity
The X-ray detector is intended for use in the electromagnetic environment specified below. The installer,
X-ray system manufacturer, or user of the X-ray detector is responsible for the usage condition of the
detector to be within such environment.
Immunity Test
Electrostatic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
transients (Burst)
IEC 61000-4-4
Transients-Surges
IEC 61000-4-5
Power frequency
magnetic field
IEC 61000-4-8
Voltage dips and
short interruptions
IEC 61000-4-11
IEC 60601 Test
Compliance
Contact: 8 kV
Contact: 8 kV
Air: 2, 4, 8, 15 kV
Air: 2, 4, 8, 15 kV
2 kV (AC mains)
2 kV (AC mains)
2 kV (DC)
2 kV (DC)
1 kV (I/O lines)
1 kV (I/O lines)
0.5, 1 kV
0.5, 1 kV
(line to line)
(line to line)
0.5, 1, 2 kV
0.5, 1, 2 kV
(line to earth)
(line to earth)
30 A/m
30 A/m
–95%/10 ms
–95%/10 ms
–100%/20 ms
–100%/20 ms
–30%/500 ms
–30%/500 ms
>–95%/5000 ms
>–95%/5000 ms
www.vareximaging.com
Electromagnetic Environment –
Guidance
Floors should be made of wood, concrete,
or ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Mains power quality should be that of a
typical commercial and/or hospital
environment.
Mains power quality should be that of a
typical commercial and/or hospital
environment.
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial and/or
hospital environment.
Mains power quality should be that of a
typical commercial or hospital
environment.
If the user of the X-ray detector requires
continued operation during power mains
interruptions, we recommend that the
X-ray detector be powered from an
un-interruptible power supply or battery.