Inspection and Maintenance
•
1.4% Hydrogen Peroxide solution:
•
Commercial option: Clorox Healthcare Hydrogen Peroxide Wipes
All chemicals listed above as approved for use can be used for cleaning and disinfecting device
surfaces. To achieve full cleaning and disinfecting results, perform the procedures in
Section 19.5.2
and
Section 19.5.3
19.5.2
Cleaning the X-ray Detector
To clean the X-ray detector:
Turn off the X-ray detector and the power and communication tethered cable, if applicable.
1
Insert the XRpad2 Protective Insert into the battery compartment before cleaning or
2
disinfecting the detector.
Locate and read the cleaning instructions specified on the product label. Follow the product
3
instructions for cleaning.
Note
If you are using a disinfectant other than those specified, we recommend
you consult a specialist for the procedure for disinfection.
Dispense wipe from container.
4
Visually inspect the device. If soil is present, use a wipe to remove soil from the device
5
surface. Use multiple wipes, as applicable, to remove all visible soil.
Take care to remove soil from seams, joints, and other difficult-to-reach areas.
6
Dispose of each wipe used for cleaning.
7
If visible residue from the cleaning product is evident after air drying, remove residue with a
8
general isopropyl alcohol wipe, and let air dry.
19.5.3
Disinfecting the X-ray Detector
To disinfect the X-ray detector:
Turn off the X-ray detector and the power and communication tethered cable, if applicable.
1
Insert the XRpad2 Protective Insert into the battery compartment before cleaning or
2
disinfecting the detector.
Locate and read the cleaning instructions specified on the product label. Follow the product
3
instructions for cleaning.
Note
If you are using a disinfectant other than those specified, we recommend
you consult a specialist for the procedure for disinfection.
66
for each cleaning product.