Commissioning
Commissioning
Installation and connection
CAUTION
• Measuring error due to an obstructed vacuum line. Prevent
• Condensate can falsify the measurement. Position the vacuum hose in
• Particles, liquids or dust may not enter the controller. Install a
• Installation and operation in areas where explosive atmospheres can
• The use of gas ballast or the operation of venting valves is only
Table top version
The table top or bench-top type controller can be installed and connected directly on top of the
work bench or on laboratory table.
The table top version is supplied with a hose nozzle. The hose nozzle should be positioned in a
way that the connected vacuum hose cannot kink.
Attachment to Hei-VAP Rotary Evaporator
The Hei-VAC Control can be attached to Hei-VAP Rotary Evaporator directly.
➜
Screw the pre-assembled metal plate that carries the vacuum valve to the back of the
evaporator with two screws:
overpressure > 1060 mbar (> 795 Torr) in the piping system
such a way that condensate cannot flow towards the controller and its
vacuum sensor. No liquid should accumulate inside the vacuum hose.
Install vacuum hoses in such a way that condensate cannot flow into
the controller.
separator or filter at the intake of the system. Appropriate filters are
for example chemically resistant, and resistant to clogging.
occur is not allowed.
permitted if thereby explosive atmospheres normally do not occur in
the interior of the equipment or, if they do occur, are likely to do so
only infrequently and for a short period. If necessary vent with inert
gas.
69