Troubleshooting
Malfunction
The unit functions but is making
too much noise.
The door has dropped or is
skewed
70 / 82
What needs to be checked?
Check if the ambient temperature
is within the permissible value
range specified in the technical
data table.
Check that thermal protection
class 3.1 or class 2 has not been
triggered.
Check that the device is not
touching any other objects.
Check if the door is properly lev-
elled.
02.2023 / Rev. --
What to you need to do?
Adjust the ambient temperature to
the value specified in this Oper-
ating Manual.
Check the set values.
Contact Customer Service.
Remove any adjacent objects
Level the unit using the adjustable
feet.
If this does not help, contact Cus-
tomer Service.