Manufacturer's Handbook and Declaration – Electromagnetic Emissions
The Chicco piston aerosol can be used in the electromagnetic environment specified below. The customer or
user of the aerosol unit must ensure it is used in such an environment
Emissions Tests
CISPR 11 Radiated/Con-
ducted Emissions
CISPR 11 Radiated/Con-
CISPR 11 Radiated/Con-
ducted Emissions
Harmonics
IEC/EN 61000-3-2
Voltage fluctuations/flickers
IEC/EN61000-3-3
Manufacturer's guide and declaration – Electromagnetic Immunity
The Chicco piston aerosol can be used in the electromagnetic environment specified below. The customer or
user of the aerosol unit must ensure it is used in such an environment.
Emissions Tests
Compliance
Electrostatic dis-
±6kV on contact
charges (ESD)
±8kV in air
IEC/EN61000-4-2
Fast/burst transients
±2kV power
IEC/EN61000-4-4
Surge
±1kV differential mode The power supply should be that typical of a commercial
IEC/EN61000-4-5
Voltage dips, brief
<5%UT for 0.5 cycle
interruptions and
variations in voltage
40%UT for 05 cycle
IEC/EN61000-4-11
70%UT for 25 cycle
<5%UT
for 5 sec
Magnetic field
3A/m
IEC/EN61000-4-8
Conducted Immunity
3Vrms 150kHz to
IEC/EN61000-4-6
80MHz (for non-life
supporting devices)
Radiated Immunity
3V/m 80MHz to
IEC/EN61000-4-3
2.5GHz (for non-life
support devices)
Note: UT is the value of the power voltage
Compliance
Group 1
Class [B]
Class [A]
[Compliant]
Guide to electromagnetic environments
Floors should be made of wood, cement or ceramic. If the
floors are covered with synthetic material, the relative humid-
ity should be now higher than 30%
The power supply should be that typical of a commercial
building or hospital.
building or hospital.
The power supply should be that typical of a commercial
building or hospital. If the aerosol user requires that the unit
operate continuously, it is recommended that it be used with
a UPS device.
The magnetic field should be that typical of a commercial
building or hospital.
13
Guide to electromagnetic environments
The Chicco piston aerosol only uses RF energy
for its internal functions. Therefore its RF emis-
sions are very low and do not cause any interfer-
ence when in proximity to electronic devices.
The Chicco piston aerosol is suited for use in
all environments, including those domestic and
those directly connected to the public electricity
distribution line that provides power to buildings
used for domestic purposes.