Herunterladen Inhalt Inhalt Diese Seite drucken

struers Labotom-3 Handbuch Seite 29

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Subject
Directives
Environment
Noise level
Electrical Data
Voltage/frequency:
3 x 200 V / 50 Hz
3 x 200-210 V / 60 Hz
3 x 220-230 V / 50 Hz
3 x 220-240 V / 60 Hz
3 x 380-420 V / 50 Hz
3 x 460-480 V / 60 Hz
Mains Cable
Voltage/ frequency
Specification
3x200V/50HZ
3x200-210/60CSA
3x220-230V/50
3x220-240/60CSA
3x380-415V/50
3x380-415/60CSA
3x460-480/60CSA
Important:
Local standards may overrule the recommendations for the main supply cable.
If necessary, please contact a qualified electrician to verify which option is suitable for the local
installation setup.
Residual
type A, 30 mA (or better) is recommended.
Current
Circuit
Breaker
Labotom-3
Instruction Manual
Specifications
Please refer to the Declaration of Conformity
Approx. 64 dB(A) running idle, at a distance of 1.0 m / 39.4" from
the machine.
Max. Load:
14.5 A
11.5 A
12.5 A
11.5 A
7.0 A
6.0 A
Cable size
3x4mm² + PE
3xAWG12 + PE
3x4mm² + PE
3xAWG12 + PE
3x4mm² + PE
3xAWG12 + PE
3xAWG12 + PE
Min.
Max.
Fuse size
Fuse size
20
25
20
25
20
25
20
25
10
20
10
20
10
20
25

Quicklinks ausblenden: