Rehalito Bedienungsanleitung / Operators Manual
Advice on Application of Electrodes
Make sure that Rehalito is turned off (no indication visible on the LCD display).
Now connect the single poled ends of the lead wires with the electrodes.
The output plugs of the electrode lead wires will be plugged into the channels A or B of the
Rehalito.
Make sure, that all plugs fit completely into the electrode sockets as well as the device sockets.
Please don't use any oil, ointment or cream, since all these have an isolating effect.
Now place the electrodes according to the instructions of your doctor or physiotherapist.
In the case of the skin gel loosing its adhesion, you can't achieve more than 6 mA - please
replace the electrodes. The same effect, however, also occurs when using a damaged lead wire.
Therefore, if the normal treatment is not possible by using new electrodes, the lead wire may be
faulty.
Electrodes which were damaged due to an improper handling may not be used.
Electrodes with a damaged lead wire isolation (e.g. visible copper wire) may not be used.
Electrodes with damaged, missing or stained aluminium coating or carbon layer, also with in-
sufficient adhesive gel may not be used.
Electrodes may not be altered (e.g. by means of scissors or other tools).
Do not pull the electrodes by force on the electrode lead wire.
Apply the electrodes on clean and health skin only.
In the case of strong bodily hair growth, the hair should be shortened with scissors.
Neither stretch the skin nor the electrodes when applying the electrodes.
To remove the electrodes from the skin, please lift the electrodes at the rim towards the middle.
Effective current densities of more than 2 mA/cm
require an increased attention of the user
2
with regard to the intensity adjustment.
If you use self-adhesive electrodes, we recommend the use of our self-adhesive electrodes with
an area of 40x80 mm.
Electrodes can contain amounts of nickel. Allergy sufferers should check this before using.
Should skin irritation occur whilst using the Rehalito, please interrupt the application and
inform your doctor - to exclude an allergic reaction.
All electrodes are made for use of one single person only.
27