Use; Restriction Of Use; Misuse Of Product; Product Life Cycle - Streifeneder KINEGEN.stream 3A2500 Gebrauchsanweisung

Polyzentrisches kniegelenk
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

7 Use

This product requires qualified and individual adaption to the patient
according to medical indication. All adjustments must be done by qualified
professional staff.

7.1 Restriction of Use

Because of the fitting and for hygienic reasons, we have to point out that the
product is to be used only by one patient and must not be given to any other
patient. The product must be used only for light sportive activities (e.g. slow
running < 7 km/h).
Attention!
We are pointing out, that this product is tested and released only
in combination with all other Streifeneder modular parts.

7.2 Misuse of Product

This product is exclusively constructed for activity levels (walky) 3 to 4 up to
maximum body weight, including carried or supported loads, as per chart,
pt. 1, technical data. The product must not be exposed to overload i.e.
impacts, shocks and vibrations.
Attention!
The approved maximum weight load of this product is limited
and depends on the component with the lowest weight limit
used in the prosthetic construction, which determines the
maximum weight load of the entire construction. The maximum
weight load means the body weight of the user including carried
and supported loads (see also the chart in technical data, pt. 1).

7.3 Product Life Cycle

The product life cycle is limited on 5 years.
36/60