Calibration of probe
OK
Calibration of probe
pH7
OK
Calibration of probe
(20)
Caution: Do not rub the probe to dry it since this causes electrostatic charge and can
result in incorrectly measured values.
Calibration of probe
pH4
OK
values
Calibration of probe
(20)
Calibration of probe
successful
in mV
Error value!
There are a number of potential causes of calibration errors:
You have used the same buffer solution twice during calibration. Correct
calibration can only be carried out with two different technical buffer solutions.
pH 7 first and then pH 4.
The measuring cable was connected incorrectly. The pH probe must be connected
to the left socket.
In the pH calibration menu item, start the mode by
pressing the OK button.
Dip the pH probe into the pH 7 buffer solution and
start by pressing the OK button.
Press OK again to trigger the countdown
After 20 seconds, the control unit shows the value of
the used buffer solution and the slope in mV. If the
calibration
is
valid,
Then take the probe from the first buffer solution and,
if possible, rinse with water and dry with a dry, lint-free
paper towel.
Now dip the probe into the second pH 4 buffer solution.
Press OK again to trigger the countdown.
After another 20 seconds, the control unit shows the
of
the
second
If the calibration was valid, continue by pressing OK
The display appears with the specification of the slope
Go back using Mode or accept using OK.
The control unit has now been successfully calibrated to
the pH probe and the data has been saved.
If the calibration process was not successful, "Error" is
displayed. Cancel using Mode or repeat using OK.
39
continue
by
pressing
buffer
solution
OK
pH
4.