Herunterladen Inhalt Inhalt Diese Seite drucken

miniland thermoadvanced pharma Bedienungsanleitung Seite 40

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 82
ENGLISH
Voltage
dips, short
interruptions
and voltage
variations
on power
supply input
lines
IEC 61000-
4-11
Power
frequency
(50/60 Hz)
magnetic
field
IEC 61000-
4-8
NOTE: UT is the a.c. mains voltage prior to application of the test
Declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment
specified below. The customer or the user of device should assure that it is
used in such an environment.
Immunity
test
0 % UT; 0.5
cycle At 0°,
45°, 90°,
135°, 180°,
225°, 270°and
315°
0 % UT; 1
cycle and
70 % UT;
25/30 cycles
Single phase:
at 0°
0 % UT;
250/300
cycles
30 A/m
level.
IEC 60601
Compliance
test level
Mains power quality
should be that of a
typical commercial
or hospital environ-
ment. If the user of
the device requires
continued operation
Not
during power mains
applicable
interruptions, it
is recommended
that the device
be powered from
an uninterruptible
power supply or a
battery.
Power frequency
magnetic fields
should be at levels
characteristic of a
30 A/m
typical location in a
typical commercial
or hospital environ-
ment.
Electromagnetic
level
40
environment -
guidance