Herunterladen Inhalt Inhalt Diese Seite drucken

Electrical Installation - Cole Parmer V-100 Serie Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für V-100 Serie:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

Thank you for purchasing this Cole-Parmer product. To
get the best performance from the equipment, and for
your personal safety, please read these instructions
carefully before use.
This equipment is designed to operate under the
following conditions:-
❖ For indoor use only
❖ Use in a well ventilated area
❖ Ambient temperature range +5°C to +40°C
❖ Altitude to 2000m
❖ Relative humidity not exceeding 80%
❖ Mains supply fluctuations not exceeding 10%
❖ Overvoltage category II IEC60364-4-443
❖ Pollution degree 2 IEC664
❖ Use with a minimum distance all round of
200mm from walls or other items
If the equipment is not used in the manner
described in this manual and with accessories
other than those recommended by the
manufacturer, the protection provided may be
impaired.

Electrical Installation

THIS EQUIPMENT MUST BE EARTHED
Before connection please ensure that the line supply
corresponds to that stated on the rating label.
Power requirements:
Model
Wattage
V-100
20W
V-200
20W
The Vortex Mixer is provided with a NEMA 5-15
plug for 120 V installations and a UK 3-pin and a
"Schuko" 2-pin plug for 230 V installations..
Choose the lead appropriate for your electrical
installation and discard the other.
Should neither lead be suitable, take the lead
with the U.K. plug and replace the plug with a
suitable alternative.
THIS OPERATION SHOULD ONLY BE UNDERTAKEN BY A
QUALIFIED ELECTRICIAN
NOTE: Refer to the equipment's rating plate to
ensure that the plug and fusing are suitable for
the voltage and wattage stated.
The wires in the power cable (230 V) are coloured
as follows:
BROWN
- HOT/LIVE
BLUE
- NEUTRAL
GREEN/YELLOW
- EARTH
The wires in the power cable (120 V) are
coloured as follows
BLACK
- HOT/LIVE
WHITE
- NEUTRAL
GREEN
- EARTH
IF IN DOUBT CONSULT A QUALIFIED ELECTRICIAN
Should the mains lead need replacing, a cable of
1mm² of harmonised code H05W-F connected to
an IEC320 plug should be used.
The appropriate mains lead should be connected
BEFORE connection to the mains supply.
Safety Advice Before Use
❖ Never move or carry the unit when in use or
connected to the mains electricity supply.
❖ DO NOT position the product so that it is
difficult to access the ON/OFF switch.
❖ DO NOT position the product so that it is
difficult to disconnect it from the mains supply
using the mains plug.
❖ The mains outlet socket used should be
located close to the equipment and readily
identifiable and accessible to users.
❖ Do not mix flammable liquids or use the
equipment in hazardous atmospheres.
❖ In the case of mains interrupting, a fault or
mechanical failure, the unit will NOT continue
to operate on removal of fault.
❖ Mechanical energy can lead to breakage of
glass vessels. Use with care.
Operation
Position the apparatus on a firm level surface,
ensuring that the IEC connector is in easy reach.
V-100 Fixed speed version (see Fig 1)
Connect the unit to the mains. Green neon
lights. To mix tube, press base firmly down onto
the rubber cap, the unit will start. Once the
contents of tube are mixed, lift clear, the unit will
stop until another tube is pressed onto the cup.
1

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

V-200 serie

Inhaltsverzeichnis