ISO/GLP-compliant Printout

Features
You can have the parameters per-
taining to the ambient weighing con-
ditions printed before (GLP header)
and after (GLP footer) the values of
a weighing series.
These parameters include:
GLP header:
– Date
– Time at beginning of measurement
– Balance manufacturer
– Balance model
– Balance serial number
– Software version number
– Identification number of the
current sampling operation
GLP footer:
– Date
– Time at end of measurement
– Field for operator signature
The record can only be output to
!
a Denver data printer 901042-1
20
Settings
§ Set the following menu codes
(see "Configuring the Balance"):
– GLP-compliant record:
menu code 5 5 2
– ISO/GLP-compliant record after
calibration/adjustment only:
menu code 7 2 2 or ISO/GLP-com-
pliant record always on: menu
code 7 2 3
– Line format for printout:
With data ID codes – 22 characters:
menu code 7 1 2
– Printout date/time:
– 24-h format: menu code 7 3 1
– 12-h format: menu code 7 3 2
No ISO/GLP-compliant record
!
is output if any of the following
settings are configured:
menu codes 6 1 3, 6 1 4 (automatic
printout) and 7 1 1
Function Keys
Press [PRINT] to output header and
first measured value.
> Header is output the first time
[PRINT] is pressed
To output header and reference data
automatically with an application
program active: Press [FUNCTION]
End application program:
End application program and output
GLP footer: Press [CF]