Guidance and manufacturer's declaration - electromagnetic emissions
The Trial Stimulator is intended for use in the electromagnetic environment specified below.
The customer or the user of the Trial Stimulator should assure that it is used in such an environment
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker
emissions IEC 61000-3-3
Table 6: Test levels for RFID Immunity testing to AIM 7351731
Frequency
134.2 kHz
13.56 MHz
433.92 MHz
860-960 MHz
2.450 GHz
Table 5 – Emission class and group
Compliance
Group 1
Class B
Not Applicable
Not Applicable
Immunity Test Level
65 A/m
5 A/m - 12 A/m
3 V/m
54V/m
54 V/m
Electromagnetic environment - guidance
The Trial Stimulator uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
The Trial Stimulator is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
12
Compliance Level
65 A/m
5 A/m - 12 A/m
3 V/m
54V/m
54 V/m