The Trial Stimulator is intended for use in the electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the Trial Stimulator can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment
(Transmitters) and the Trial Stimulator as recommended below, according to the maximum power output of the
communications equipment.
Rated maximum output
power of transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation application to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
Table 4 – Recommended separation distances
Recommended separation distances between portable and
mobile RF communications equipment and the Trial Stimulator
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.2 √ P
0.12
0.38
1.2
3.8
12
80 MHz to 800 MHz
d = 1.2 √ P
0.12
0.38
1.2
3.8
12
11
800 MHz to 2.7 GHz
d = 2.3 √ P
0.23
0.73
2.3
7.3
23