Equipmentrequirements
All manuals and user guides at all-guides.com
• An accurate digital multimeter (accurate to 0.0 mV and ±0.1 μA)
Outputcalibrationprocedure
Whenthe
Action/Description
displayshows
Allow fifteen minutes of powered operation (to reach a stable temperature).
Put the instrument in setup mode and scroll through the main menu
COPn
Press
Calibrate Outputs?
COPy
Measure the output current
Press
I=0
(Use the keypad to adjust the value as required)
I=20
Press
Measure the output voltage
E=0
Press
E=10
Press
Instrument will save values and return to normal operation (using the new values).
Save
Note: Do not remove the power while the save message is on display.
22
*
.
*
. The output current should be 0.00 mA
*
. The output current should be 20.00 mA
*
. The output voltage should be 0.00 V
*
. The output voltage should be 10.00 V
ì
Å
Toggle
Increase Decrease
Increase Decrease
Increase Decrease
Increase Decrease
ENT
Start
Accept
Accept
Accept
Accept