What to do in the event of measuring errors or operating errors
People suffering from cardiac arrhythmia, vascular constrictions, arteriosclerosis in the extremities and diabetics
should consult their doctor before performing blood pressure self-monitoring because in such cases the blood pres-
sure levels may be different.
Error
ERR 1: Measuring error
ERR 2: Inflation error
ERR 3: Deflation error
ERR 300
GB-50
Cause
Arm or body movement during measurement.
The cuff is not sitting properly.
Talking during measurement.
No pressure is being built up in the system.
Deflation rate too high or too low.
A fault has been detected.
Inflation pressure above 300 mmHg
Battery error
Remedy
Repeat measurement after 3-5 minutes. Do
not move your arm and do not talk. If the error
occurs again, contact customer service.
Repeat measurement and if the error occurs
again, call the service telephone number.
If the error occurs again, call the service
telephone number.
Repeat measurement after 3-5 minutes. Do
not move your arm and do not talk.
Repeat measurement after 3-5 minutes.
Check batteries