MAINTENANCE
EN
CHECK IT
Check the blue water filter and
change it if needed.
EACH YEAR

Preventive maintenance

Send your device to your authorized EMS repair center.

TROUBLESHOOTING

Handpiece clogged
1
2
3
AIR
H
O
2
Use Easy Clean
Place the Easy Clean
Then take a syringe
Dry both sides of handpiece
into the handpiece.
filled with water.
with compressed air.
Service needed
Missing cord
Stop using your device.
Check your cord connection.
Batteries low
Overheating
Change pedal batteries.
Wait for cooling.
If error repeats, call service.