Herunterladen Inhalt Inhalt Diese Seite drucken

General Specification - TensCare perfect mama+ Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Output will
Open circuit cut-out
not increase
operating.
above zero
Lead not connected to
body or faulty/damaged.
(Lead wires can break at the bend where they leave the machine giving
no, or intermittent, output).
Sudden
If you disconnect and re-
change in
connect a few minutes
sensation
later, the signal will feel
quite a lot stronger.
If the above review has failed to resolve your problem, or to report unexpected
operation or events, or to provide feedback call TensCare or your local supplier or
distributor (address on back cover) for advice.
Contact TensCare customer service on +44 (0) 1372 723 434. Our staff are trained to
assist you with most issues you may have experienced, without the need to send your
product in for service.
European Medical Device Regulation requires that any serious incident that has
occurred in relation to this device should be reported to the manufacturer and the
competent authority in your country. This can be found at:
https://ec.europa.eu/docsroom/documents/3683/attachments/1/translations/en/renditi
ons/pdf

18. GENERAL SPECIFICATION

Waveform
Amplitude
(over 500 Ohm load)
Output plug
Channels
Batteries
Weight
Dimensions
Safety Classification
Ensure machine is attached to your body
correctly.
Try using the second lead wire.
Purchase replacement if necessary.
Always return strength to zero after
disconnecting the leads or electrode pads.
Symmetrical Bi-Phasic Rectangular
70 mA zero to peak, 50 steps in A & B, 60 steps in C
+/- 10%
Fully shielded: touch proof
Dual channel
2 x AA alkaline (two AA batteries)
75 g without batteries
120 x 60 x 20 mm
Internal power source.
Designed for continuous use.
No special moisture protection.