Maintenance
9.2.3

Battery Check

To ensure the battery performance, you must perform regular maintenance for
the battery. The method is as follows:
1. Ensure that the device is not being used for a patient, and power off the device.
2. Connect the device to the AC power supply through the power adapter to
recharge the battery for over 4 hours.
3. Disconnect the device from the AC power supply after the indicator becomes
green.
4. Power on the device for the device to operate with the display screen
brightness and light source brightness being set to the default levels. Record
the start time.
5. Make the device operate continuously until it stops when the battery is empty.
Record the end time.
 If the operation lasts for 0.8 to 1.2 hours, the battery is close to the end of
its service life.
 If the operation lasts for less than 0.8 hours, the battery has reached the end
of its service life. In this case, the battery must be replaced. You are
advised to contact the distributor for battery replacement.
After battery status check is completed, recharge the battery for next use.
33/ 46