MD 9930 Thermal camera
Camera supports three temperature ranges: Screening, -20°C ~ 150°C and
0°C ~ 350°C.
Screening range covers temperatures 32°C ~ 42°C and is used exclusively for body
temperature measurement and fever screening on people. It provides the highest
accuracy of 0.5°C in this range.
NOTE: Alarms are only valid if set to 36.0°C or higher when using Screening range.
The camera may beep continuously if the alarm is set lower.
For general measurements of temperature in the overlap of other two ranges, the
lower range provides best results.
•
In main menu, navigate to Range menu using UP and DOWN buttons.
•
Open by pressing SELECT.
•
Use UP and DOWN buttons to highlight, then SELECT to choose the
range.
•
The change will be applied when camera recalibrates, which happens right
after the menu is closed and may take a few seconds to complete.
Settings
Settings contain general options, localization, and factory reset.
In main menu, press UP and DOWN buttons to highlight "Settings".
Use SELECT button to open the submenu. Options in submenu are: Language
selection, Temperature unit, Factory reset, Set date / time, Information data,
Alignment, Auto power off, and Volume. These options are available in two pages.
The second page is accessible by using DOWN button after bottom option on first
page is selected.
Press SELECT button to select an item and change its settings or CLOSE button to
exit.
Figure 8: Settings menu, pages 1 and 2
Language
•
Navigate to Language submenu in the Settings and open it.
•
Choose a language using UP, DOWN, LEFT and RIGHT buttons.
20
Settings Menu