Declaration – electromagnetic emissions and immunity
for equipment and systems that are not life-supporting and are specified for use only in a shielded location
Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
-5% UT(95% dip in UT)
for 0.5 cycle, -40 % UT
(60% dip in UT) for 5 cycles,
-70% UT(30 % dip in UT) for
25 cycles, -5% UT
(95% dip in UT) for 5 sec
30 A/m 50 Hz or 60 Hz
N/A
The main power quality should be that of a typical
commercial or hospital environment. If the user of the
equipment or system requires continued operation during
power main interruptions, it is recommended that the
equipment or system be powered from an uninterruptible
power supply or a battery.
30 A/m 50 Hz and 60 Hz
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.