Technical Data

GB
Instrument
Light source:
Wavelength accuracy
Photometric accuracy*
Photometric resolution
Power supply
Operating time
Auto-OFF
Display
Storage
Serial Interface
Bluetooth
Modul:
®
Time
Calibration
Dimensions
Weight
Ambient conditions
Waterproof
CE
*measured with standard solutions
To ensure maximum accuracy of test results, always use the reagent systems supplied by
the instrument manufacturer.
54
Technical Data
double
wavelength, automatic wavelength selection,
direct reading colorimeter
LEDs,
interference filters (IF) and photosensor in transparent
cell chamber. Wavelength specifications of the IF:
530 nm
= 5 nm
560 nm
= 5 nm
± 1 nm
3% FS (T = 20° C – 25° C)
0.01 A
4 batteries (AAA/ LR 03)
17hr operating time or 5000 measurements in constant
use when display backlight and Bluetooth
are off.
automatic switch off
10 minutes after last keypress
backlit LCD (on keypress)
internal ring memory for 125 data sets
Bluetooth
4.0 for the transfer of stored measurement data
®
Specifications:
Modul: BLE113-A
Bluetooth
4.0 LE
®
FCC ID: QOQBT113
IC: 5123A-BGTBLE113
real time clock und date
user and factory calibration
resetting to factory calibration possible
155 x 75 x 35 mm (LxWxH)
approx. 260 g (incl. batteries)
temperature: 5 – 40 °C
rel. humidity: 30–90 % (non-condensing)
floating; as defined in IP 68 (1 hour at 0.1 meter)
Certificate for Declaration of CE-Conformity