Measurement Principle And Process - Knauer K-7400 Benutzerhandbuch

Halbmikro-osmometer
Vorschau ausblenden Andere Handbücher für K-7400:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

10
Fig. 5

Measurement Principle and Process

Fig. 6
Putting down the measuring head without vial
The osmotic pressure in a solution is proportional the freezing point
depression of that solution. An aqueous solution, with an osmotic
pressure corresponding to an ideal 1-molal solution, freezes at -1.858 °C.
An aqueous solution with this freezing point has a concentration of
1 Osm/kg.
To perform a measurement first the sample solution is cooled down
without stirring. Even pure water can be super-cooled down to about -5
through -8°C without freezing. The freezing is initiated by an
automatically start of the vibrator at a set temperature. The temperature
reaches the freezing point, 0°C in case of pure water.
°C
10
0
Super-cooling
-10
0
1
Temperature-time-curve for pure water
The freezing point of a solution is below 0°C. The freezing point
depression T is a measure for the osmolality of a solution, which can be
red directly from the display given in mOsm/kg.
Description of the Osmometer K-7400
Freezing point
2