Using the EuroOsmo® 7400 Software
Fig. 13
Fig. 14
Measuring window with the temperature sequence of a run
Setup area of the main window
Setup
Operator
Communication Port Select the COM port to which your osmometer is
Working Parameters
Monitoring view
Measurement view
Freezing Level
Cooling Limit
Units
Area for entering general settings for measurement
sequences and corresponding reports.
Enter the operator's name
connected.
The entered value defines the scale of the monitor
by its final value in minutes. The temperature
pattern is shown continuously for the defined
period.
The entered value defines the scale of the
measurement window by its final value in minutes.
The initiation point „freeze" can be set within the
range from -13.0 to -4.0 °C. Any decimal value will
be rounded in steps of 0.5°.
The minimum temperature can be set within the
range from -18.0 to -12.0 °C. Any decimal value
will be rounded in steps of 0.5°.
Select the units valid for the measurement
sequence. The choice is mOsmol, °C, °F, and %. If
°F is selected the values of the freezing level and
cooling limit are automatically changed to °F.
By clicking on the Apply button the entered values
are sended to the osmometer. If any value is out of
the allowed range a corresponding message will
21