Herunterladen Inhalt Inhalt Diese Seite drucken

LR Suprasorb CNP P3 Gebrauchsanweisung Seite 150

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Error message
Status
A signal sounds.
Therapy continues
to run in the
background.
A signal sounds.
Therapy is
interrupted.
A signal sounds.
A signal sounds.
Therapy is
interrupted.
Possible cause
Remedy
Low battery level.
If the battery level is low, please connect
the device to the mains power supply
using the supplied power supply.
If the battery level is empty, please
connect the device to the mains power
supply immediately using the supplied
power supply.
Therapy has not
You now have the option of starting
been started.
the therapy or switching the device off.
A simple trick often helps in such cases:
Switch the device off and on again, then
check whether it is functioning normally.
If this does not help, please contact
your CNP specialist or the L&R hotline
(see Section 11).