CAUTION
As soon as the exudate pouch reaches maximum capacity, the Suprasorb
the message "Exudate pouch is full". Please then proceed as follows:
1. Confirm the message by pressing the green tick.
3. Switch the Suprasorb
4. Detach the exudate pouch tube from the wound drainage system connector.
Exudate pouch
• Ensure that the exudate pouch for the Suprasorb
regularly. When the exudate pouch is full, the integrated overflow prevention
system is triggered. This discontinues the therapy.
• Switch the device to Pause mode when replacing the exudate pouch.
2. Close the tube clamps on the exudate pouch tube
and the wound drainage system to ensure that the
negative pressure in the wound is maintained.
CNP P3 off
(Pause mode -> see Section 5.4).
®
5. Detach the container system from the remaining
unit by first pushing down and then to the side,
as shown in the diagram.
CNP P3 is replaced
®
CNP P3 displays
®