LEAKAGE TEST
After connection to medical gas pipeline system terminal unit or to medical
regulator, visually check possible leakages:
low pressure regulator inlet connection to terminal unit or to medical
regulator,
inlet connector to low pressure regulator body,
user outlet.
If any leakage is detected, use the procedure in chapter 6.3 and return
the product to GCE for service.

FUNCTIONAL TEST

Check that there is a gas flow at each setting (for instance, by listening
for the sound of gas flow or checking presence of bubbles in an
humidifier).
Reset flow control knob on ZERO position.
6.2. CONNECTION & USE OF OUTLET
LIST OF RECOGNIZED ACCESSORIES
To be connected to the flow outlet:
humidifiers, breathing masks or cannulas, gas savers, nebulizers.
Before connecting any accessory or medical device to the low pressure
regulator, always check that it is fully compatible with the product
connection features & the product performances.
FLOW OUTLET CONNECTION
When connecting any accessory to the flow outlet make sure that it is not
connected to the patient before operating the product.
Ensure the hose/humidifier is compatible with the flow outlet feature.
Push the hose onto the regulator flow outlet/outlet of humidifier.
Ensure the hose/humidifier is well engages.
FLOW SETTING
Set medical gas flow control knob on one of the available flow rates by
turning the flow control knob in the requested position.
Always ensure that the flow control knob has engaged and not placed
between two settings otherwise the flow selector will not deliver correct
flow of medical gas.
Do not try to apply an excessive torque on the flow control knob when it
stops on the maximum flow position or in zero position.
Only user and doctor is accountable for method of using at specific
patient and for assessment of flow value.