Accessing a parameter
1.
2.
3.
4.
5.
Changing a Parameter Value
Notes:
• Security must be disabled to enable programming: set P000 to the Unlocked Value stored
in P069. (A remote master can still change configuration, if P799 is set to allow this.)
• Invalid entries will be rejected or limited.
• CLEAR
1.
2.
Parameter Reset to Factory Default
1.
2.

Master Reset (P999)

Returns all parameters except P000, P069, and P838 to default settings. (The learned TVT curve
is not lost.)
1.
2.
3.
4.
5.
A5E32106153
Downloaded from
www.Manualslib.com
Press PROGRAM
then DISPLAY
PROGRAM mode.
Either use the ARROW keys
different parameter, or:
Press DISPLAY
to open the Parameter Number
field. (The current parameter value remains visible.)
Key in the desired parameter number followed by
ENTER
.
For example: press
The LCD displays the new parameter number and
value.
can be used to clear the field.
Key in the new value.
Press ENTER
to set the value.
Scroll to the parameter or enter its address.
Press CLEAR
then ENTER
Press PROGRAM
then DISPLAY
,
Press DISPLAY
to open parameter fields.
Key in 999.
Press CLEAR
then ENTER
reset. The LCD displays C.ALL.
Reset complete. (Reset takes several seconds to complete.)
SITRANS Probe LR – QUICK START MANUAL
manuals search engine
, to activate
to scroll to a
.
. The value returns to the default setting.
to activate PROGRAM mode.
, to Clear All and initiate
PROGRAM Icon
Parameter Number
Parameter Value
current value
Page EN-11