Reusable cuff: Wipe the cuff with a damp cloth and clean
with soap or disinfectant. The cuffs have been tested to
withstand the following recommended disinfectants: Buraton
liquid, Cidex®, Ethanol 70%, Isopropyl-acohol 70%, Mikrozid®
and Sporicidin®. After disinfection, the cuff cover should be
rinsed with water and left to air dry.
Single patient cuffs: The single patient cuff is to be used on
one and the same patient for the duration of their treatment.
The cuffs can be personalized. Do not clean or disinfect the
single patient cuffs! Dispose of the cuff properly after use!
Warnings:
1. When the cuff is not connected to the gauge, always use
a tube stopper to prevent accidental penetration of liquid.
The penetration of liquid into the tube may cause damage
to the gauge.
2. Periodically inspect and replace damaged or deteriorated
cuffs.
3. Use only the appropriate size cuffs. The arterial pointer
must be in the marked area, otherwise the measurement
readings may be incorrect.
4. Single-patient cuffs are not intended for use on more than
one patient.
5. Do not immerse the device in liquid to clean it.
6. Do not sterilize the device!
7. The blood pressure measuring device must be handled
carefully and must not be exposed to pressure, sharp
objects or any other mechanical stress.
8. The pointer on the gauge must point within the tolerance
range of the zero position when there is no pressure in the
cuff.
9. Always let all of the pressure out of the cuff before storing
the instrument.
10. Protect the surface of the stethoscope diaphragm from
damage.
11. Never inflate to more than 300 mmHg.
12. When replacing or repairing parts, please ensure that only
original KaWe replacement parts are used.
www.kawemed.com
13. Make sure that the tube system is airtight.
14. It is recommended that the device be handed over to the
manufacturer or an authorized service center at least every
2 years for proper metrological control according to the
Medical Devices Operator Ordinance.
No changes to the device are allowed!

Technical specifications:

Measurement range: 0-300 mmHg
Measurement accuracy: ± 3 mmHg
Storage/operating conditions:
Gauge:
Storage conditions: - 30 °C to +70 °C
Operating conditions: + 10 °C to +40 °C
Relative humidity: 15% - 85%
Cuff:
Storage conditions: - 20 °C to +55 °C
Relative humidity: 15% - 90%
Operating conditions: + 10 °C to +40 °C
Sea level: from -70 m to +1700 m
Warranty:
If used properly and in accordance with the instructions
for use, the warranty period is two years from the date
of purchase. If used properly and in accordance with the
instructions for use, the product life is at least 10 years. If you
have any questions or need repairs, please contact your dealer.