Safety Measures for Use in Hazardous Areas

2.1.1 Starting Up and Starting Up after Service

The DMA 35
N
labelled with the ATEX registration number (see appendix C).
The intrinsically safe DMA 35
G EEx ib IIC T5. All instructions given in the certificate of conformity have to be
followed (see appendix C).
By labelling the intrinsically safe DMA 35
Anton Paar GmbH confirms that the instrument conforms with all documents
submitted for receiving the certificate of conformity. Therefore, do not subject the
intrinsically safe DMA 35
Repaired instruments may be operated only if the repair was carried out by
Anton Paar GmbH. The type plate according to ATEX still confirms that the
intrinsically safe DMA 35
the certificate of conformity.

2.1.2 Battery Exchange

Use only batteries of the same type and in the same level of charge.
When exchanging the batteries, always check that the small parting plate is
inside the battery compartment chamber and properly fixed.
After the battery exchange (outside the hazardous area), close the battery cover
and fix it with the Phillips screw.
Check the function of the instrument: Turn the instrument on and check the
display.

2.1.3 Infrared/RS 232 Interface Port

Never use the infrared/RS 232 interface port in hazardous areas.
B46IB01
is suitable for use in hazardous areas only if the instrument is
can be used according to ATEX marking
N
to alterations of any kind.
N
conforms with all documents submitted for receiving
N
with the type plate according to ATEX,
N
II 2
9