Medical Electrical Equipment needs special precautions regarding EMC and needs to be
installed and put into service according to the EMC information provided in the following.
Portable and mobile RF communication equipment (e.g. cell phones) can affect Medical
Electrical Equipment.
The use of accessories and cables other than those specified may result in increased emissions
or decreased immunity of the unit.
Guidance and manufacturer's declaration – electromagnetic emissions
The A&D unit is intended for use in the electromagnetic environment specified below.
The customer or the user of the A&D unit should assure that it is used in such an
environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Recommended separation distances between portable and mobile RF
communications equipment and the A&D unit
The A&D unit is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the A&D unit can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the A&D unit as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where p is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Compliance
Electromagnetic environment – guidance
The A&D unit uses RF energy only for its
internal function. Therefore, its RF emissions
Group 1
are very low and are not likely to cause any
interference in nearby electronic equipment.
The A&D unit is suitable for use in all
Class B
establishments, including domestic
establishments and those directly connected to
Class A
the public low-voltage power supply network
that supplies buildings used for domestic
purposes.
N.A.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
d = 1.2 P
0.12
0.38
1.2
3.8
12
English 19
m
800 MHz to 2.5 GHz
d = 1.2 P
0.12
0.38
1.2
3.8
12
d = 2.3 P
0.23
0.73
2.3
7.3
23

Werbung