Verfügbare Sprachen

Verfügbare Sprachen

schwa-medico
Setting parameter for program no. 1.0, 1.1, 1.2, 1.3
(standard stimulation)
Select the program with the
Press the
E
key for 4 seconds to enter the parameter modifi-
cation mode. This is indicated by an acoustic signal and the
word "SET" flashes on the display.
Press ▲▼ key of channel 1 or 3 to modify the intensity
between 1 Hz and 120 Hz.
Press ▲▼ key of channel 2 or 4 to modify the pulse width
between 50 µs and 300 µs.
Press the
E
key to save the changes.
Press ▲▼ key of channel 1 or 3 to modify the time between 10 min and 90 min.
4 seconds after pressing the last key, the device automatically returns to the stimu-
lation mode. This is indicated by an acoustic signal.
The stimulation can then be started by increasing the intensity with the ▲ keys of
each connected channel.
Setting parameter for program no. 3.0, 3.1, 3.2 (HAN stimulation)
Select the program with the
Press the
E
key for 4 seconds to enter the parameter modification mode. This is
indicated by an acoustic signal and the word "SET" flashes on the display.
Press ▲▼ key of channel 1 or 3 to modify the intensity between 1 Hz and 120 Hz.
Press ▲▼ key of channel 2 or 4 to modify the pulse width between 50 µs and 300 µs.
Press the
key to save the changes.
E
Press ▲▼ key of channel 1 or 3 to modify the time between 10 min and 90 min.
Press the
key to save the settings of the first frequency. You can now set the
E
second parameters which will alternate every 3 seconds with the first parameters
you have selected.
Press ▲▼ key of channel 1 or 3 to modify the intensity between 1 Hz and 120 Hz.
Press ▲▼ key of channel 2 or 4 to modify the pulse width between 50 µs and 300 µs.
4 seconds after pressing the last key, the device automatically returns to the stimu-
lation mode. This is indicated by an acoustic signal.
The stimulation can then be started by increasing the intensity with the ▲▼ keys
of each connected channel.

Factory settings

Press the key for 4 seconds to retrieve the factory settings and delete the chan-
ged parameters in the AS SUPER 4 digital. This is indicated by an acoustic signal.