Tab. 2) Guidance and manufacturer's declaration - electromagnetic immunity
The Device is intended for use in the electromagnetic environment specified below. The customer or the user of the Device should assure
that it is used in such an environment
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
NOTE UT is the a.c. mains voltage prior to application of the test level.
100275_MYO-TENS_USER MANUAL_REV.00.indd 48
line(s) to line(s): ±1 kV
line(s) to line(s): ±1 kV
line(s) to earth: ±2 kV
100 kHz repetition frequency
100 kHz repetition frequency
5% UT
Not applicable
(>95% dip in UT)
for 0.5 cycle
40% UT
(60% dip in UT)
for 5 cycle
70% UT
(30% dip in UT)
for 25 cycle
<5% UT
(>95% dip in UT)
for 5s
30 A/m
30 A/m
50Hz/60Hz
50Hz/60Hz
Mains power quality should be that of a typical commer-
cial or hospital environment.
Mains power quality should be that of a typical commer-
cial or hospital environment.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.